Long-term outcomes of percutaneous closure of ventricular septal defects in children using different devices: A single centre experience from Egypt

Abstract

Background

The feasibility of percutaneous closure ventricular septal defects (VSD) in children has been previously proven. However, data on long-term outcomes are limited. We aim to evaluate the long-term outcome of our experience with percutaneous closure of VSD using various occluders.

Methods

Retrospective institutional analysis of children who underwent transcatheter closure of perimembranous and muscular VSDs between September 2012 and February 2020. Patient demographics, procedural, and long-term follow-up data were comprehensively analyzed. Patients who lost to follow-up within two years post-procedure were excluded.

Results

We identified 75 patients (54.7% males) with a median of 66 months (IQR, 46–96). The closure success rate at one year was 95.7%. Complete heart block was detected in two patients early post-procedure and resolved with steroids. The VSDs were perimembranous (52%), muscular (33.33%), and residual (14.67%). Implanted devices were Pfm Nit-Occlud LeˆVSD Coil (42.7%), HyperionTM VSD Muscular Occluder (28%), Amplatzer VSD muscular occluder (10.7%), Amplatzer Duct Occluder (14.7%), Occlutech Muscular VSD Occluder (2.7%), and Amplatzer Duct Occluder II (1.3%). No new arrhythmia or valve regurgitation was detected after two years post-procedure. Persisted complications on long-term follow-up included: residual shunting in 3(4%), mild tricuspid regurgitation in 2(2.7%), and aortic regurgitation in 2(2.7%), with one immediate post-catheterization mild aortic regurgitation worsened during follow-up, requiring surgical repair of VSD three years after device implantation. No deaths were reported.

Conclusion

Long-term outcomes of pediatric transcatheter VSD closure using different devices are satisfactory. Post-procedural adverse events are limited, but long-term surveillance is necessary to monitor their progression.

Peer Review reports

Introduction

Ventricular septal defect (VSD) is the most common congenital heart disease, accounting for 30% of all congenital cardiac malformations [1]. Closing VSD is sometimes needed because spontaneous closure may not occur with subsequent potential complications. Surgical closure remains the mainstay of treatment, especially for large defects. However, some defects are better approachable percutaneously [2].

The first VSD transcatheter closure was reported by Lock et al. in 1988, using a Rashkind double-umbrella device [3]. Since then, we witnessed huge advances in device technology and imaging modalities, and device closure has become widely adopted to close hemodynamically significant native or residual defects, thereby avoiding cardiopulmonary bypass and lengthy hospital stays [4,5,6,7,8,9].

The immediate and midterm results of percutaneous closure of VSD have been extensively reported, with well-established procedure safety and efficacy [10,11,12,13,14,15,16]. However, data on longer-term outcomes using variable occluders is still limited. Therefore, we aim to present our institutional experience with transcatheter VSD closure and evaluate the long-term outcomes of the procedure.

Materials and methods

We performed a retrospective clinical data review of pediatric patients (≤ 18 years old) who underwent device closure of perimembranous and muscular VSD at our institution. The Institutional review board (IRB) of Mansoura University, Faculty of Medicine, Egypt, approved the study. Informed consent was obtained from the patient’s legal guardians. Patients who lost to follow-up within two years post-procedural were excluded from this study.

Indications for VSD closure at our institution were hemodynamically significant left-to-right ventricular shunt or a history of infective endocarditis related to the VSD. The hemodynamically significant VSD was defined by the presence of at least 3 of the following criteria: (1) overt heart failure, not improving with appropriate medication, (2) failure to thrive, predominantly due to the hemodynamic effects of the VSD, (3) recurrent respiratory infections, (4) Increased cardiothoracic ratio on chest X-ray, (5) left ventricular (LV) end-diastolic z-score on echocardiogram, indexed to body surface area of ≥ 2.0, and (5) estimated pulmonary to systemic blood flow ratio of > 1.5 at cardiac catheterization [6, 17].

Cases not eligible for closure were: (1) large defects ≥ 10 mm, (2) bidirectional or predominantly right to left shunt on color Doppler, (3) aortic valve prolapse or aortic regurgitation more than trivial, and (4) doubly committed VSDs or complex cardiac anatomy requiring surgical intervention.

Procedure

All procedures were performed under general anesthesia, and fluoroscopy with transesophageal or transthoracic echocardiographic guidance. Intravenous antibiotic prophylaxis (cefotaxime or cefuroxime 50 mg/kg, up to 1 g) was administered before the procedure. Short venous and arterial femoral access was obtained, preferably on the right side. Intravenous Heparin (100 IU/kg, up to 5000 IU) was given to keep activated clotting time above 200s. First, routine right and left heart catheterizations were done to evaluate the pulmonary-to-systemic flow ratio (Qp/Qs). Then, left ventriculography and aortography, in lateral and 30° right anterior oblique/15° caudal projections, were performed to delineate the VSD anatomy and evaluate the presence of aortic valve prolapse or regurgitation.

A 4 or 5 French partly cutoff pigtail or Judkins right catheter was advanced to the left ventricle or ventricular outflow tract. Then, a retrograde or antegrade approach was used for VSD closure [18].

Available devices during the study period were: Amplatzer Duct Occluder I (ADO I), (Abbott, USA); Amplatzer Duct Occluder II (ADO II), (Abbott, USA); Amplatzer Muscular VSD Occluder (AMO), (Abbott, USA); Nit-Occlud Leˆ VSD coil (PFM Medical, Germany); Hyperion™ VSD Muscular Occluder (Comed B.V., Netherlands/Lepu MT Company, China); and Occlutech Muscular VSD Occluder (Occlutech GmbH, Germany).

Device selection depends on the defect morphology, location, relation to the aortic valve, the diameter of the defect, and device availability in the armamentarium. For native perimembranous and residual VSDs, ADO I was used in the presence of significant-sized aneurysms and a subaortic rim ≥ 5 mm. The ADO I right disc diameter was selected to be 2 mm more than the smallest VSD diameter. ADO II was used in defects with no aneurysmal tissue, diameter < 5 mm, and subaortic rim > 3 mm. The chosen ADO II waist diameter was 1 mm more than the smallest VSD diameter. Nit-Occlud Leˆ VSD coil was used in aneurysmal VSD with a diameter < 8 mm and subaortic rim ≥ 3 mm. The selected coil had a distal end at least twice the minimal VSD diameter on the right ventricular side and equal to or 1–2 mm greater than the left ventricular opening of VSD. Hyperion device was used in some cases with a defect to aortic valve distance >4 mm as the length of the device’s left retention skirt is 4 mm. For muscular VSDs, one of the muscular occluders was used (AMO, Hyperion, Occlutech), according to the device availability of an appropriate size in the catheterization lab. However, AMO was available throughout the duration when our patients had their VSD closure, but Hyperion use in our patients was started in 2018 and Occlutech in early 2020. The size of the muscular occluder depends on the device’s waist diameter, which was chosen to be 1–2 mm larger than the VSD.

Through the long sheath, the device was deployed under fluoroscopic guidance. Before release, left ventriculography was performed 10 min after device implantation to verify the occluder position, residual shunt, and valvular condition. After the procedure, oral aspirin 3–5 mg/kg, up to 100 mg daily, was prescribed for six months in all patients.

Successful closure was defined as the absence or trivial residual shunting. Trivial shunting was defined as a minimal signal in color Doppler without a complete signal on pulsed/continuous Doppler [19].

Post-procedural follow-up

Follow-up evaluation was performed on day one post-procedure, three and six months, then every 12 months. Device position, VSD residual shunt, and valvular condition were evaluated by transthoracic echocardiography. Holter monitoring was performed when an abnormality was suspected on 12-lead electrocardiography (ECG).

Last follow-up visit

The time elapsed from the intervention was documented. Two-dimensional, color flow, Doppler, and three-dimensional echocardiography were performed to document residual flow, device position, aortic valve morphology/function, and adverse events. The presence of heart block or abnormal rhythm in the last follow-up was looked at.

Complications

Life-threatening adverse events or those requiring surgical management were defined as serious complications. Adverse events requiring medical or transcatheter intervention were defined as major complications. Complications that did not require specific management were termed minor complications [19].

Statistical analysis

Statistical analyses were performed using SPSS, version 25 (IBM, Armonk, NY, USA). Categorical variables were reported as frequency and percentage, while continuous variables were represented as median with interquartile range (IQR). Statistical analyses for continuous variables were conducted using Kruskal-Wallis test, while Chi-square test and Fisher’s exact test were used for categorical variables as appropriate. A p-value < 0.05 was considered statistically significant. All reported P values are two-sided.

Results

From September 2012 to February 2020, 96 patients had an attempted closure of VSD. Implantation was initially achieved in 94/96 patients with a technical success of 97.9%. Of the 94 cases, most cases were treated successfully with a single procedure; nevertheless, four patients required a second device insertion within the same procedure; one due to immediate embolization after release, another due to partial device herniation before release, and the other two because of significant residual shunt verified by ventriculography before device release. After the device release, the closure immediate success rate was 79/94 (84%), which increased at six months to 91. 5%, and reached 90/94 (95.7%) at one year of follow-up, with no further increase on subsequent follow-up. Long-term follow-up was possible in 75/94 (79.8%) patients with inserted devices (see Fig. 1). Eighteen cases were lost to follow-up (none had immediate complication), with one death unrelated to the procedure (traffic accident).

Fig. 1
figure 1

Chart demonstrating the study’s cohort of patients. n: Number of patients, m: months, y: year

Table 1 demonstrates the demographic and cardiac lesion characteristics of the included patients. The median follow-up duration (IQR) was 66 (46–96) months with a minimum follow-up of 28 months, and the highest follow-up duration was 115 months (9.6 years). The study included 75 patients; 54.7% were males. The median (IQR) age of cases at intervention was 9 (7–11) years (ranging from 3 to 17 years), and the median weight (IQR) was 30 (21–43) kg (ranging from 12 to 77 kg). Most patients had native perimembranous VSD in 39 (52%), followed by muscular VSD in 25 (33.3%), and residual VSD post-cardiac surgeries in 11 (14.7%). Of the residual VSD cases, 6 had right bundle branch block (RBBB) pre-catheterization. A concomitant cardiac lesion was present in 19/75 (25.3%); one patient had a PDA closure transcatheter in the same sitting (see Fig. 2), and another had pulmonary valvuloplasty simultaneously with VSD closure.

Table 1 Demographics and cardiac lesion characteristics of the included patients in the long-term follow-up
Fig. 2
figure 2

(A) LV angiography (60° left anterior oblique with 20° cranial) to determine the location and size of VSD (red arrow). (B) Retrograde approach for implanting Hyperion device across VSD (yellow arrow). (C) Antegrade approach for implanting Amplatzer muscular occluder device across VSD (yellow arrows) with LV angiography before the occluder released, LV angiography showed no residual shunt and satisfying position and shape of the occluder. (D) LV angiography after Amplatzer muscular occluder release with no residual shunt. (E) Aortic angiography demonstrating small elongated PDA (red arrow) after closing VSD with ADO I (yellow arrow). (F) Two devices are seen in place; Vascular plug II closing PDA (red arrow) and ADO I across VSD (yellow arrow)

Procedural and device data of the included patients in the study are demonstrated in Table 2. Most transcatheter VSD occluders were implanted through an antegrade approach in 53.3%, while the retrograde approach was used in 46.7% (see Fig. 2). Moreover, the most used echocardiographic guidance was transesophageal in 85.3% of cases. Pfm Nit-Occlud Leˆ VSD Coil was the most used device in 32 (42.7%) cases, especially for perimembranous VSD 26 (66.7%), followed by HyperionTM VSD Muscular Occluder 21(28%), particularly for the muscular VSD type.

Table 2 Procedural and device data of the included patients in the study

Details of early post-procedural and long-term mortality and morbidity are documented in Table 3. Early Minor complications were detected in 28.7% of cases. One of the major complications encountered was device embolization. It occurred in two patients; the first had a Hyperion device embolized immediately upon release, but it was snared and replaced with a larger device. The other was an ADO I device embolized one day after the procedure, which was retrieved transcatheter with surgical closure later based on the parents’ request (see Fig. 3). Access-related limb ischemia requiring systemic heparin infusion occurred in one child.

Table 3 Short- / long-term procedure-and device-related morbidity and mortality
Fig. 3
figure 3

Early complications of VSD device closure (A) LV angiography demonstrating coil in place across subaortic perimembranous VSD with mild residual shunt (B) Post-procedure transthoracic echocardiography showing residual shunt (C) Hand injection in right common iliac artery demonstrating embolized ADO I into the bifurcation of the aorta without causing obstruction (D) ADO I embolized Device was successfully snared. Yellow Arrow is pointing at the device

Complete atrioventricular block (cAVB) was detected in 2 (2.1%) patients with perimembranous VSD closed with ADO I. A transient block was detected on the first day after the catheterization in one patient and on the third day in the other; both resolved on prednisolone by the fifth- and seventh days post-procedure without temporary pacing. None of our cases had serious events and no short-term mortality.

On long-term follow-up, two cases had persistence of the new tricuspid regurgitation, three resolved on follow-up, and three patients had persisted residual shunt. However, none of them showed hemodynamic consequences. One case with mild aortic regurgitation early post-procedure progressed to severe regurgitation by the end of the second year, causing heart failure in the early third year of follow-up. Surgical removal of the device and patch closure of VSD was performed. The surgeon reported that the valve’s right coronary cusp was trapped underneath the device. However, the valve did not require any intervention, as the regurgitation was less upon device removal. No new emerging complication was documented on long-term follow-up and no mortalities. Echocardiographic images of some long-term encountered complications are demonstrated in Fig. 4.

Fig. 4
figure 4

Long-term follow-up using transthoracic echocardiography: (A, B) Two-dimensional four-chamber view and left parasternal long-axis view (in rightward and inferior angulation toward right hip) demonstrating mild tricuspid regurgitation caused by Hyperion muscular device (yellow Arrow) used to close a high muscular outlet VSD (C), Three-dimensional parasternal long-axis view of ADO I is seen slightly protruding in the LVOT causing minimal aortic regurgitation and (D) Long parasternal view demonstrating aortic regurgitation caused by a device in closing the perimembranous VSD.

Table 4 compares different devices used for VSD closures regarding demographic, device size, complications, and long-term follow-up. Although no statistically significant results were found regarding the age, sex, and weight of patients at the time of device implantation, significant differences were detected in the size of the device and follow-up duration (P-value = 0.0001). The largest device sizes were used for Muscular Amplatzer, followed by Hyperion. The longest follow-up duration was for ductal occluders and Amplatzer muscular devices, and the least was for Occlutech. Primarily it could be attributed to Occlutech being a recently released device used in the most recent patients. The immediate residual flow was the highest following coil in 28% of cases; however, this rate declined to 3.1% on follow–up. The highest incidence of new tricuspid regurgitation was detected with Hyperion devices in 9.5% of cases, while the ADOI was the most frequently implicated in new post-procedural aortic regurgitation 2 (18.2%).

Table 4 Comparison between devices regarding demographic, device size, complication, and long-term follow-up

Discussion

The percutaneous closure of VSDs has become a standard practice utilizing different occluding devices, enabling an expansion in managing various types and morphologies of VSD. The current work thoroughly investigated the outcome of transcatheter closure of VSD using six types of devices over a long duration of follow-up, reaching up to 9.5 years in some cases.

Residual shunt

The technical success in our series was achieved in 97.9% of patients, the immediate success rate was 84%, and the maximum achieved success rate was 95.7% at one year without further complete closure beyond a year of follow-up. The highest rate for closure was 100% for AMO, and the lowest rate for immediate closure was for the Pfm Nit-occlud LeˆVSD coil, which improved on the follow-up to a closure rate consistent with other devices. In agreement with our results, a success rate of 91% was detected in the Rahmath et al. cohort after 54.5 months of follow-up [17]. A lower ultimate closure rate than ours was reported by Bergmann et al. (86.2%), with the highest rates on using Amplatzer membranous and muscular occluders (93–95%) and the lowest with Nit-Occlud LeˆVSD coil at (61%); with a new small shunt detected at 40 weeks of gestation in a pregnant patient three years after using the coil [20]. In EUREVECO Registry, Nit-occlud LeˆVSD coil’s immediate closure rate was reported to be as low as 50%, but increased to 97% after one year [14]. Nevertheless, Walavalkar et al. found that the probability of device failure was not associated with device type, as they did not detect a significant difference in failure rate between muscular and ductal devices [21]. For residual post-operative VSDs, we had successful closure in 10/11 (90.9%) patients; only one had a small residual. Comparable results were reported by Taha et al. with successful device closure of 18 residual VSDs either post-operative or post-catheterization with no reported complication or mortality after a mean follow-up of 23.3 months [22].

Tricuspid regurgitation

New-onset trivial/mild tricuspid regurgitation was diagnosed early after device implantation in 5.3% of our cohort without progression, with a resolution of the regurgitation in two patients. In contrast, Rahmath et al. documented a higher rate of new tricuspid regurgitation immediately post-procedural in 18 (40%) with a resolution of mild regurgitation in seven patients, while moderate regurgitation persisted at 54.5 months of follow-up [17]. On the contrary, a lower rate of mild degree tricuspid regurgitation was detected by Mandal et al. in 1.1% of their patients [10]. However, progression from mild degree post-procedure to severe tricuspid regurgitation causing heart failure was reported three years following transcatheter VSD closure due to coaptation failure of the valve leaflets as a result of septal leaflet entrapment by the device [23]. Similarly, severe tricuspid regurgitation requiring surgical repair with coil removal was reported in two other cases of perimembranous VSD closed with Nit-Occlud LeˆVSD Coil [19]. Impingement of the device on the septal leaflet of the tricuspid valve and rupture of the chordae tendineae are possible etiologies of regurgitation in rare situations. Therefore, the abnormal origin of the tricuspid valve main chordae tendineae from a perimembranous VSD was screened for and was considered an exclusion criterion for transcatheter closure in some centers [24].

Aortic regurgitation

Another VSD occluders-related complication is aortic valve regurgitation. Our study documented new-onset regurgitation in 3 (3.2%) patients early post-procedure. Only one case progressed to a severe degree, requiring surgery. In contrast, Han et al. reported aortic regurgitation in 0.1% of their cohort; four patients had severe regurgitation 9–12 years post-VSD transcatheter closure requiring replacement [25]. Another report documented aortic regurgitation after four years with no hemodynamic consequence in 1 (0.6%) case [10]. On the contrary, a high incidence of new immediate trivial/mild aortic regurgitation was reported in 5/45 (11.1%) patients in another study, increased to 13.3% on long-term follow-up for 54.5 months; however, none exceeded a mild degree of regurgitation [17]. Similarly, Walavalkar documented new immediate aortic regurgitation in 15% of patients, reduced to 5% at a median follow-up of 246 days using 15 different devices [21]. New-onset or worsening aortic regurgitation was the primary cause of unplanned surgery after VSD transcatheter closure in children [26]. Aortic regurgitation could result from iatrogenic cusp injury or perforation during the placement or retrieval of a device [27]. Furthermore, it could be the occluder effect on the aortic valve as a large occluder may get too close to the aortic valve, and the memory alloy of an occluder may gradually erode the surface in contact, causing delayed aortic regurgitation [28, 29].

Arrhythmia

In the current work, the rate of early post-procedure arrhythmia was 4.3%; only two patients with perimembranous VSD closed using ADO I had early cAVB resolved on steroid therapy without pacing. However, no atrioventricular block (AVB) was detected on prolonged tracking of cases. Unlike our data, Zhao et al. noted high arrhythmia rates, reaching 24.1% early following transcatheter closure; 77.8% reverted to sinus rhythm during 35.5 months of follow-up. On logistic regression, they found a significant relation between arrhythmia and long fluoroscopy time when using eccentric or large devices [30]. Another study on perimembranous VSD found that 25.5% of cases had early arrhythmia post-device insertion with 2.7% serious arrhythmias, including second & third AVB and left bundle branch block (LBBB), with reported late onset in seven patients (6 months to 8.3 years later) [31]. Another series reported that 8.5% of patients had conduction abnormalities; with transient cAVB occurring in two patients, junctional rhythm (27 patients), RBBB (3 patients), and LBBB (2 patients) [31]. Bergmann et al. 4/109 (3.7%) found variable arrhythmias after one year of follow-up, including one case of SVT [20].

The incidence of cAVB associated with transcatheter closure of perimembranous VSD was reported to be 1–6%, depending on different clinical experiences and occluder selection [32]. A meta-analysis of percutaneous device closure of pmVSD revealed that the incidence was 1.1% [33]. The cAVB rate was 0.7%, as reported by Bergmann et al., with no other reported arrhythmia within six years of follow-up for six different devices [20]. However, Li et al. reported a higher rate of early conduction abnormalities after perimembranous VSD device 19/79 (24.05%); 11 cases of incomplete RBBB (6 resolved on 35.3 ± 17.4 months follow-up) and 5 cases of complete RBBB (one resolved on long-term follow-up), and reversible cAVB; two of them received temporary pacemaker implantation. These patients recovered one, six, and nine days later with no new reported cases on long-term follow-up for 3–5 years [34]. Nevertheless, a new AVB could develop years after device insertion; as Xie et al. reported, a cAVB case was diagnosed 2.5 years post perimembranous VSD closure using a modified double-disc VSD occlude [35]. Also, Bai et al. reported that three patients developed AVB beyond one year of VSD device, one of these developed 5.3 years post-procedure [36]. It was suggested that AVB occurring immediately after the procedure might result from direct mechanical compression by the device or inflammatory edema of the membranous interventricular septum near the AV node and conduction branches. However, late block after weeks or months post-procedural may result from a localized inflammatory reaction caused by the device that can result in extensive scar tissue and cartilaginous metaplasia of the surrounding myocardium.). Another proposed mechanism for late cases of AVB is device flattening [37, 38]. Therefore, the early block tends to respond to steroid therapy, while the late block requires management with a permanent pacemaker. A single report described a rare recurrent cAVB 42 months after the VSD transcatheter closure complicated with transient cAVB one week after the procedure [39].

The proximity of the membranous septum to the septal leaflet of the tricuspid valve on the VSD right ventricular side and the aortic valve on the left ventricular side, with its inferoposterior margin closely related to the bundle of His and bundle branches, is considered a challenging anatomy. Therefore, successful closure of the VSD transcatheter requires complete anatomic delineation of the defect, careful VSD sizing, and careful determination of the relation to surrounding cardiac structures for proper device selection to minimize the risk of complications. On the one hand, undersized devices are associated with device embolization and residual shunt. On the other hand, oversized devices may damage adjacent structures, causing cAVB or injury of aortic or tricuspid valves [40].

Infective endocarditis

No cases of infective endocarditis were detected in the current series through the follow-up duration. However, rare cases were reported in the literature. In one case, early VSD device-related infective endocarditis was reported in a patient ten days after a Nit-Occlud LeˆVSD coil insertion [41]. A case of pulmonary valve endocarditis was reported by Carminati et al. a few days post-implantation of two Amplatzer VSD muscular occluders [42]. Another case was diagnosed with Kingella kingae endocarditis four months after transcatheter closure of perimembranous VSD using a Nit-Occlud Leˆ VSD coil [43]. Late-onset infective endocarditis could occur following device closure if associated with post-procedure aortic regurgitation, as reported by Tang et al. 11 years post-VSD closure using symmetric double-disk occluder (SHAMA) [44].

Mortality

No device-related mortality for percutaneous closure of VSD was documented in the current series post-procedure or on the long-term follow-up that was extended up to 9.5 years. Our finding is consistent with several published studies reporting no deaths on short- or long-term follow-ups for percutaneously implanted devices [14, 20, 45, 46]. Limited cases of early procedure-related mortalities were documented, such as Carminati et al. report of one intraprocedural mortality in a case during second AMO implantation, resulting in a mortality rate of 0.2% [42]. Moreover,  there is a report of one patient died five days following catheterization due to intracranial bleeding [21]. Furthermore, Jiang et al. reported two (0.3%) deaths; one developed subarachnoid hemorrhage due to cerebral vascular malformation, whereas the other was arrested secondary to cAVB 40 days post-procedure. However, they did not document additional mortalities on a long-term follow-up duration of 46 months [47].

Although the current study proves the favorable outcome of VSD percutaneous closure after several years of follow-up, the study design has inherent limitations. One of the primary limitations was the retrospective design of the study. Furthermore, the results might not represent those encountered in different centers as it is a single-center experience. Moreover, the patients’ sample size was insufficient for some devices for a solid statistical analysis. Therefore, future prospectively designed studies with larger sample sizes and the involvement of multiple centers are fundamental to confirm the long-term outcome of variable devices used in VSD percutaneous closure.

Conclusion

In conclusion, according to our experience, transcatheter closure of VSD is safe in pediatric patients with no related mortality in immediate or long-term surveillance. Long-term safety is comparable among different devices. Although post-procedural adverse events were limited, extended follow-up is required for cases with early reported complications for the limited potential of progression to significant morbidity, especially the aortic valve regurgitation.

Data Availability

The data are not publicly available because they contain information that could compromise the privacy of research participants in this study. Data is available from the corresponding author upon reasonable request.

References

  1. Hoffman JI. Incidence of congenital heart disease: I. postnatal incidence. Pediatr Cardiol. 1995;16:103–13. https://doi.org/10.1007/BF00801907.

    Article  CAS  PubMed  Google Scholar 

  2. Zhu D, Tao K, An Q, Luo S, Gan C, Lin K. Perventricular device closure of residual muscular ventricular septal defects after repair of complex congenital heart defects in pediatric patients. Tex Heart Inst J. 2013;40(5):534–40. PMID: 24391313.

    CAS  PubMed  PubMed Central  Google Scholar 

  3. Lock JE, Block PC, McKay RG, Baim DS, Keane JF. Transcatheter closure of ventricular septal defects. Circulation. 1988;78(2):361–8. https://doi.org/10.1161/01.cir.78.2.361.

    Article  CAS  PubMed  Google Scholar 

  4. Knauth AL, Lock JE, Perry SB, McElhinney DB, Gauvreau K, Landzberg MJ, et al. Transcatheter device closure of congenital and postoperative residual ventricular septal defects. Circulation. 2004;110(5):501–7. https://doi.org/10.1161/01.CIR.0000137116.12176.A6.

    Article  PubMed  Google Scholar 

  5. Dua JS, Carminati M, Lucente M, Piazza L, Chessa M, Negura D, et al. Transcatheter closure of postsurgical residual ventricular septal defects: early and mid-term results. Catheter Cardiovasc Interv. 2010;75(2):246–55. https://doi.org/10.1002/ccd.22262.

    Article  PubMed  Google Scholar 

  6. Feltes TF, Bacha E, Beekman RH 3rd, Cheatham JP, Feinstein JA, Gomes AS, et al. Indications for cardiac catheterization and intervention in pediatric cardiac disease: a scientific statement from the American Heart Association. Circulation. 2011;123(22):2607–52. https://doi.org/10.1161/CIR.0b013e31821b1f10.

    Article  PubMed  Google Scholar 

  7. Kouakou NYN, Song J, Huh J, Kang IS. The experience of transcatheter closure of postoperative ventricular septal defect after total correction. J Cardiothorac Surg. 2019;14(1):104. https://doi.org/10.1186/s13019-019-0933-8.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Liu H, Lu FX, Zhou J, Yan F, Qian SC, Li XY, et al. Minimally invasive perventricular versus open surgical ventricular septal defect closure in infants and children: a randomised clinical trial. Heart. 2018;104(24):2035–43. https://doi.org/10.1136/heartjnl-2017-312793.

    Article  CAS  PubMed  Google Scholar 

  9. El-Kadeem S, El Nemr S, El Amrousy D, Zoair A. Comparison of transcatheter versus surgical closure of perimembranous ventricular septal defect in pediatric patients: a systematic review and meta-analysis. J Saudi Heart Assoc. 2019;31(4):188–97. https://doi.org/10.1016/j.jsha.2019.06.002.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Mandal KD, Su D, Pang Y. Long-term outcome of Transcatheter device Closure of Perimembranous Ventricular Septal defects. Front Pediatr. 2018;6:128. https://doi.org/10.3389/fped.2018.00128.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Koneti NR, Sreeram N, Penumatsa RR, Arramraj SK, Karunakar V, Trieschmann U. Transcatheter retrograde closure of perimembranous ventricular septal defects in children with the Amplatzer duct occlude II device. J Am Coll Cardiol. 2012;60:2421–2. https://doi.org/10.1016/j.jacc.2012.08.1004.

    Article  PubMed  Google Scholar 

  12. Ghaderian M, Merajie M, Mortezaeian H, Aarabi Moghadam MY, Shah Mohammadi A. Mid-term follow-up of the Transcatheter Closure of Perimembranous Ventricular Septal Defects in Children using the Amplatzer. J Tehran Heart Cent. 2015;10(4):182–7. PMID: 26985206.

    PubMed  PubMed Central  Google Scholar 

  13. Ghaderian M, Merajie M, Mortezaeian H, Aarabi M, Mohammad Y, Shah Mohammadi A. Efficacy and safety of using Amplatzer Ductal Occluder for Transcatheter Closure of Perimembranous ventricular septal defect in Pediatrics. Iran J Pediatr. 2015;25(2):e386. https://doi.org/10.5812/ijp.386.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Haas NA, Kock L, Bertram H, Boekenkamp R, De Wolf D, Ditkivskyy I, et al. Interventional VSD-Closure with the Nit-Occlud® Lê VSD-Coil in 110 patients: early and Midterm results of the EUREVECO-Registry. Pediatr Cardiol. 2017;38(2):215–27. https://doi.org/10.1007/s00246-016-1502-8.

    Article  PubMed  Google Scholar 

  15. El Shedoudy S, El-Doklah E. Mid-term results of transcatheter closure of ventricular septal defect using nit-occlud Lê ventricular septal defect coil, single-center experience. J Saudi Heart Assoc. 2019;31(2):78–87. https://doi.org/10.1016/j.jsha.2018.11.002.

    Article  PubMed  Google Scholar 

  16. Yang J, Yang L, Yu S, Liu J, Zuo J, Chen W, et al. Transcatheter versus surgical closure of perimembranous ventricular septal defects in children: a randomized controlled trial. J Am Coll Cardiol. 2014;63(12):1159–68. https://doi.org/10.1016/j.jacc.2014.01.008.

    Article  PubMed  Google Scholar 

  17. Rahmath MR, Numan M, Dilawar M. Medium to long-term echo follow-up after ventricular septal defect device closure. Asian Cardiovasc Thorac Ann. 2016;24(5):422–7. https://doi.org/10.1177/0218492316645746.

    Article  PubMed  Google Scholar 

  18. Haddad RN, Daou L, Saliba Z. Device Closure of Perimembranous ventricular septal defect: choosing between Amplatzer Occluders. Front Pediatr. 2019;7:300. https://doi.org/10.3389/fped.2019.00300.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Kozlik-Feldmann R, Lorber A, Sievert H, Ewert P, Jux C, Müller GC, et al. Long-term outcome of perimembranous VSD closure using the Nit-Occlud® Lê VSD coil system. Clin Res Cardiol. 2021;110(3):382–90. https://doi.org/10.1007/s00392-020-01750-6.

    Article  PubMed  Google Scholar 

  20. Bergmann M, Germann CP, Nordmeyer J, Peters B, Berger F, Schubert S. Short- and long-term Outcome after Interventional VSD Closure: a single-center experience in Pediatric and adult patients. Pediatr Cardiol. 2021;42(1):78–8. https://doi.org/10.1007/s00246-020-02456-2.

    Article  CAS  PubMed  Google Scholar 

  21. Walavalkar V, Maiya S, Pujar S, Ramachandra P, Siddaiah S, Spronck B, et al. Percutaneous device Closure of congenital isolated ventricular septal defects: a Single-Center Retrospective Database Study Amongst 412 cases. Pediatr Cardiol. 2020;41(3):591–8. https://doi.org/10.1007/s00246-020-02315-0.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Taha FA, Alnozha F, Amoudi O, Almutairi M, Abuelatta R. Transcatheter Closure of residual and iatrogenic ventricular septal defects: Tertiary Center Experience and Outcome. Pediatr Cardiol. 2022;43(2):308–23. https://doi.org/10.1007/s00246-021-02722-x.

    Article  PubMed  Google Scholar 

  23. Sarkislali K, Kalangos A. Late tricuspid regurgitation after Percutaneous Transcatheter Closure of ventricular septal defect: an Educational Presentation. Braz J Cardiovasc Surg. 2021;36(2):253–6. https://doi.org/10.21470/1678-9741-2020-0172.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Wang L, Cao S, Li J, Yang L, Liu Y, Ren J, et al. Transcatheter closure of congenital perimembranous ventricular septal defect in children using symmetric occluders: an 8-year multiinstitutional experience. Ann Thorac Surg. 2012;94:592–8. https://doi.org/10.1016/j.athoracsur.2012.03.067.

    Article  PubMed  Google Scholar 

  25. Han Y, Li H, Zhu H, Sun G, Yin Q, Gu C. Aortic regurgitation after closure of ventricular septal defect by transcatheter device: the long-term complication. Cardiol Young. 2020;30(4):577–9. https://doi.org/10.1017/S1047951120000414.

    Article  PubMed  Google Scholar 

  26. Yang P, Wu Z, Liu Z, Zhang J, Zhou H, Ji X. Unplanned surgery after Transcatheter Closure of ventricular septal defect in children: causes and risk factors. Front Pediatr. 2021;9:772138. https://doi.org/10.3389/fped.2021.772138.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Haddad RN, Sawan EB, Saliba Z. Word of caution: severe aortic valve injury linked to retrograde closure of perimembranous ventricular septal defects. J Card Surg. 2022;37(6):1753–8. https://doi.org/10.1111/jocs.16441.

    Article  PubMed  Google Scholar 

  28. Tuluca A, Omer S, Cornwell L, Blaustein A, Kar B, Weldon S, Bakaeen FG. Aortic valve leaflet entrapment by a percutaneous closure device. Ann Thorac Surg. 2014;98(1):e23–5. https://doi.org/10.1016/j.athoracsur.2014.02.080.

    Article  PubMed  Google Scholar 

  29. Negi S, Kumar A, Podder S, Mishra AK. Percutaneous ventricular septal defect closure causing acute aortic regurgitation. Egypt J Cardiothorac Anesth. 2018;12:12–5. https://doi.org/10.4103/ejca.ejca_20_17.

    Article  Google Scholar 

  30. Zhao LJ, Han B, Zhang JJ, Yi YC, Jiang DD, Lyu JL. Postprocedural outcomes and risk factors for Arrhythmias following Transcatheter Closure of congenital Perimembranous Ventricular Septal defect: a single-center Retrospective Study. Chin Med J (Engl). 2017;130(5):516–21. https://doi.org/10.4103/0366-6999.200551.

    Article  PubMed  Google Scholar 

  31. Shah JH, Saraiya SP, Nikam TS, Jha MJ. Transcatheter device Closure of Perimembranous ventricular septal defect in Pediatric Patients: long-term outcomes. Heart Views. 2020;21(1):17–21. https://doi.org/10.4103/HEARTVIEWS.HEARTVIEWS_13_19.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Zhou Y, Qin Y, Zhao X, Lang X, Zhu N, Bai Y et al. The Impact of Short or Long Transcatheter Occluder Waist Lengths on Postprocedure Complete Atrioventricular Block: A Retrospective Study. J Invasive Cardiol. 2015; 27(11):E231-5. PMID: 26524207.

  33. Santhanam H, Yang L, Chen Z, Tai BC, Rajgor DD, Quek SC. A meta-analysis of transcatheter device closure of perimembranous ventricular septal defect. Int J Cardiol. 2018;254:75–83. https://doi.org/10.1016/j.ijcard.2017.12.011.

    Article  PubMed  Google Scholar 

  34. Li P, Zhao XX, Zheng X, Qin YW. Arrhythmias after transcatheter closure of perimembranous ventricular septal defects with a modified double-disk occluder: early and long-term results. Heart Vessels. 2012;27(4):405–10. https://doi.org/10.1007/s00380-011-0155-z.

    Article  CAS  PubMed  Google Scholar 

  35. Xie L, Zhang H, Zhang R, Xiao T. Management of late-onset complete atrioventricular Block Post Transcatheter Closure of Perimembranous Ventricular Septal defects. Front Pediatr. 2020;7:545. https://doi.org/10.3389/fped.2019.00545.

    Article  PubMed  PubMed Central  Google Scholar 

  36. Bai Y, Liu J, Qin YW, Wu H, Zhao XX. Percutaneous Closure of Perimembranous Ventricular Septal defect with modified double-disk occluder: what is the outcome at 10-year Follow-up? Congenit Heart Dis. 2016;11(1):45–51. https://doi.org/10.1111/chd.12284.

    Article  PubMed  Google Scholar 

  37. Walsh MA, Bialkowski J, Szkutnik M, Pawelec-Wojtalik M, Bobkowski W, Walsh KP. Atrioventricular block after transcatheter closure of perimembranous ventricular septal defects. Heart. 2006;92(9):1295–7. https://doi.org/10.1136/hrt.2005.084988.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  38. Yalonetsky S, Lorber A. Late high degree atrioventricular block after percutaneous closure of a perimembranous ventricular septal defect. Cardiol Young. 2009;19:298–300. https://doi.org/10.1017/S1047951109003813.

    Article  PubMed  Google Scholar 

  39. Yang R, Sheng YH, Cao KJ, et al. Late recurrent high degree atrioventricular block after percutaneous closure of a perimembranous ventricular septal defect. Chin Med J (Engl). 2011;124(19):3198–200.

    PubMed  Google Scholar 

  40. Shahanavaz S, Winlaw DS, Opotowsky AR. What is blocking transcatheter ventricular septal defect Closure? J Am Heart Assoc. 2022;11(7):e024963. https://doi.org/10.1161/JAHA.122.024963.

    Article  PubMed  PubMed Central  Google Scholar 

  41. El-Sisi AM, Menaissy YM, Bekheet SA. Infective endocarditis following coil occlusion of perimembranous ventricular septal defect with the Nit-Occlud((®)) Le device. Ann Pediatr Cardiol. 2016;9(1):59–61. https://doi.org/10.4103/0974-2069.171355.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Carminati M, Butera G, Chessa M, De Giovanni J, Fisher G, Gewillig M, et al. Transcatheter closure of congenital ventricular septal defects: results of the European Registry. Eur Heart J. 2007;28(19):2361–8. https://doi.org/10.1093/eurheartj/ehm314.

    Article  PubMed  Google Scholar 

  43. Kassis I, Shachor-Meyouhas Y, Khatib I, Khoury A, Le TP, Lorber A. Kingella endocarditis after closure of ventricular septal defect with a transcatheter device. Pediatr Infect Dis J. 2012;31(1):105–6. https://doi.org/10.1097/INF.0b013e31823c3ec1.

    Article  PubMed  Google Scholar 

  44. Tang C, Zhou K, Hua Y, Wang C. Very late-onset endocarditis complicated by non-significant aortic regurgitation after device closure of perimembranous ventricular septal defect. Med (Baltim). 2020;99(19):e20120. https://doi.org/10.1097/MD.0000000000020120.

    Article  Google Scholar 

  45. Butera G, Carminati M, Chessa M, Piazza L, Micheletti A, Negura DG, et al. Transcatheter closure of perimembranous ventricular septal defects: early and long-term results. J Am Coll Cardiol. 2007;50(12):1189–95. https://doi.org/10.1016/j.jacc.2007.03.068.

    Article  PubMed  Google Scholar 

  46. Khoshhal SQ, Al-Mutairi MB, Alnajjar AA, Morsy MM, Salem SS, Al-Muhaya M, et al. Transcatheter device closure of ventricular septal defects in children: a retrospective study at a single cardiac center. Ann Saudi Med. 2020;40(5):396–402. https://doi.org/10.5144/0256-4947.2020.396.

    Article  PubMed  PubMed Central  Google Scholar 

  47. Jiang D, Han B, Zhao L, Yi Y, Yi Y, Zhang J, Fan Y, et al. Transcatheter device Closure of Perimembranous and Intracristal Ventricular Septal defects in children: medium- and long-term results. J Am Heart Assoc. 2021;10(11):e020417. https://doi.org/10.1161/JAHA.120.020417.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

None.

Funding

This research did not receive any specific grant.

Open access funding provided by The Science, Technology & Innovation Funding Authority (STDF) in cooperation with The Egyptian Knowledge Bank (EKB).

Author information

Authors and Affiliations

Authors

Contributions

HE: conceptualized the design of the study and supervised the research, MH: Helped in formulating the study design, and patient follow-up using echocardiography, GA: patient follow-up using echocardiography, AB: helped in data collection, SR: data collection, statistical analysis, wrote the initial manuscript. All authors: participated in the catheterization procedure, revised, edited and accepted the manuscript.

Corresponding author

Correspondence to Shaimaa Rakha.

Ethics declarations

Ethics approval and consent to participate

The authors assert that all procedures contributing to this work comply with the ethical standards of the relevant national guidelines on human experimentation and with the Helsinki Declaration of 1975, as revised in 2008, and the research has been approved by the institutional research board (IRB) of Faculty of Medicine, Mansoura University. Furthermore, informed consent was obtained from the guardians of the included cases.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions