Skip to main content

Validity and reliability of electroacoustic probe for diagnosis of developmental dysplasia of the hip

Abstract

Background

Sound transmission is used in the diagnosis of hip dysplasia since the end of the 80’s. Aim of this study is to quantify the validity and reliability of electroacoustic probe for the diagnosis of hip dysplasia in neonates.

Methods

Diagnostic study included neonates aged 4–28 days, whose parents signed an informed consent. The probe was used three times for comparative sound transmission and with extension/flexion; hip ultrasound was performed with Graf technique as gold standard. Kappa was determined for intraobserver and interobserver reliability; validity was calculated with sensitivity, specificity, and predictive values.

Results

100 neonates were included. For the comparative sound transmission, 0.80 and 0.81 Kappa were obtained for the intraobserver and interobserver respectively; with extension/flexion, Kappa 0.98 and 0.95 were obtained for the intraobserver and interobserver respectively. With comparative sound transmission, 44.8%, 97.7%, 76.5% and 91.3% for sensitivity, specificity, positive and negative predictive values, respectively; with extension/flexion test, the sensitivity, specificity, positive and negative predictive values: 82.8%, 99.4%, 96.0%, and 97.1%, respectively.

Conclusion

The electroacoustic probe is moderate valid and reliable for the diagnosis of developmental dysplasia of the hip.

Trial registration

Open Science framework https://osf.io/kpf5s/?view_only=0a9682c6w1c842ad8e1d9a66e8dcf038

Peer Review reports

Background

Developmental dysplasia of the hip (DDH) is a range of hip disorders ranging from slight incongruence between the articular surfaces of the ilium and femur to the displacement of the femoral head out of the acetabulum [1]. In Mexico, it is considered that 1% of newborns have hip dysplasia and 75% of macrosomic infants have ultrasound evidence of alterations in the hip; although the evolution of hip dislocation occurs only in 1: 7000 live births [2]. In the United States, DDH is estimated at 1 in 100 infants in the form of instability and in 1 out of every 1000 newborns in the form of hip dislocation [3].

The health professional who takes care of children, is facing a challenge as it is considered, according to Fernandez, that 73% of affected children are diagnosed by parents during the second six months of life [4], although it seems that up to 95% of cases have gone unnoticed by health professionals [1], overshadowing the forecast, and therefore reaching to more invasive treatment and higher chances of disability in the function of the lower extremities.

The clinical diagnosis is made by performing clinical maneuvers such as Ortolani, Barlow, limitation of abduction, asymmetric folds, Galeazzi, and Piston, but these only detect hip subluxation or dislocation [1, 5, 6] and it is intended to achieve an early diagnose using the physical properties of the bone for sound transmission [59].

Stone et al., described the use of a tuning fork and stethoscope to diagnose DDH,[7] using comparative test sound transmission (CTST) and comparative sound transmission with extension / flexion (STE/F). Padilla et al., applied both tests in children under 2 years with stethoscope and tuning fork [5], and in neonates with a tuning fork [6, 8], which reported greater validity than the usual clinical maneuvers.

The CTST and STE/F with stethoscope and tuning fork had greater validity than the usual clinical maneuvers [5, 6, 8], but this is a subjective test since a good auditory acuity and good training is required.

Padilla et al., evaluated a device based on the transmission of sound with high validity and reliability [9], where the sound wave is propagated through the bone, from the patella to the pubic symphysis, where it is perceived by a receiver which communicates with an amplifier and converts the sound received into digits and subsequently confirming the diagnosis with an ultrasonography hips. With results previously reported by Padilla et al., in 2014 [9], an electroacoustic probe was designed, this is an electromagnetic device that is capable of producing sinusoidal signals at a frequency that is required in a range of 1 Hz to 400 kHz and can be an integer or decimal. It also includes a sound wave generated by an electroacoustic amplifier pitch; this is transmitted through the lower end of the patella to the pubic symphysis. A touch screen shows the sound transmission recorded in graphic or digital form. The receiver of the electroacoustic probe is placed on the pubic symphysis and the perceived sound wave is transformed which is then presented in decibels on the screen of the device, allowing to perform similar tests to the tuning fork and stethoscope tests, with the difference of this being an objective test (Fig. 1).

Fig. 1
figure 1

Electroacoustic probe

With sound transmission tests, tuning fork and stethoscope, bone radar (Mexican Pat. N°337,887) or electroacoustic probe, for healthy hips the transmission sound is the same and equal in both hips, but if there is an immature or dysplasic hip, the sound is lower because the contact between components of the hip is minor; with the sound transmission test with extension/flexion, when the extremity is flexed the contact between components of the hip is higher if the hip is immature or dysplasic, and the sound transmission increases compared with flexion.

The aim of the study was to identify the reliability and validity of the electroacoustic probe for DDH diagnosis in neonates of Celaya, Guanajuato.

Methods

The protocol was approved by the Bioethics Committee of the Division of Health Sciences and Engineering Campus Celaya Salvatierra, University of Guanajuato.

It is a diagnostic test study based on the community. It was held in Celaya, Guanajuato between January and December 2014.

Mothers of newborns from three public and four private hospitals of the city, were invited to participate in the study, performing it on the facilities of the University of Guanajuato.

Selection of participants

Inclusion criteria

Neonates of 4–28 days old, whose parents agreed in writing that their child may participate.

Exclusion criteria

Newborns with rigid, embryological hip dislocation.

Variables

Gender, age, area of ​​residence, birth weight and height, weight and height when beginning study were measured.

The CTST and STE/F were applied with the electroacoustic probe. For CTST, the newborn was placed supine with legs extended; the electroacoustic tuning fork was placed on the left kneecap and the receiver on the pubic symphysis; on the digital display of the amplifier “on” is pressed and the sound transmission lasts 5 s, whose wave is picked up by the receiver and the results in decibels appear on the screen. The electroacoustic tuning fork is placed on the opposite patella and the same procedure was performed. If the sound is lower in any of the sides, this is an indicative of an alteration in the hip.

For STE/F, the newborn was placed in a supine position with the pelvic limbs extended. The electroacoustic tuning fork is placed on the left kneecap and the receiver on the pubic symphysis, pressing “on” generates a sound wave for 5 s, which is picked up by the receiver and the results in decibels are shown on the screen. The hip is flexed at 90 ° and the measurement procedure is repeated. If the sound increases while bending, this indicates an altered hip (Fig. 1).

The newborns subsequently underwent hip ultrasonography using Graf’s method, which was applied with a portable ultrasound transducer, Honda MS2000. The static and dynamic tests were performed and the angles alpha and beta were measured for both tests on both hips. The following criteria was taken for the diagnosis of DDH [10, 11]:

I Graf angle α > 60 ° and angle β <55 °, healthy hip

Graf II 44–59 ° angle α and angle β 55 ° -77, physiological immaturity

Graf III and IV angle α <43 ° and β angle > 77 °, subluxation or dislocation.

Procedures

Invitations to participate were distributed to parents in vaccination units; in private hospitals, mothers who gave birth were also invited. Those who attended the University of Guanajuato, were given an information sheet for parents, and the formulated questions were answered. Later they were asked to sign an informed consent. Those neonates who made it had their height and weight measured and their parents were asked about the date of birth, gender, birth height and weight, and area of ​​residence. Thereafter the CTST and STE/F tests were applied to the newborns, twice with the electroacoustic probe by an investigator and a third time by a different investigator. The newborn immediately underwent hip ultrasound using Graf’s technique; the ultrasonographer was blinded to the results of the sound tests.

Sample size

Expecting a sensitivity of 85% with a prevalence for DDH of 10%, the minimum sample size is 11 neonates with a 95% of precision and 90% of power (3.1 Epidat, 2005, Xunta de Galicia and PAHO)

Statistic analysis

Descriptive statistics were used for the study variables.

For intraobserver reliability, Kappa was calculated by comparing measurements 1 and 2 and interobserver Kappa by comparing measurements 1 and 3.

For the validity, the sensitivity, specificity, and predictive values were calculated for the test of sound transmission with electroacoustic probe using Graf hip ultrasound technique as gold standard.

The statistical analysis was performed in 13.0® STATA (Stata Corp., College Station, TX, USA).

Results

The sample consisted of 200 hips from 100 newborns. Female neonates predominated (64%), newborns living in urban areas (72%), 16% reported having a family history of DDH, parents being the most frequently reported as affected (4%) and 9% other relatives (cousins, uncles or grandparents) (Table 1).

Table 1 Sociodemographics categorical characteristics of newborns, Celaya, 2014 (n = 100)

The quantitative characteristics of the infants were: age range 4–28, average 14.7 ± 7.9 days; birth weight was 1.9 to 4.2 with a mean of 3160.3 ± 426.9 g; height at birth was 44–55 with an average of 49.8 ± 2.1 cm; weight at the beginning of the study was 2270–5100 with an average of 3538.7 ± 586.3 g; height at the beginning of study was 47–60 with an average of 51.9 ± 2.7 cm.

Reliability for CTST and STE/F is shown in Table 2, excellent intraobserver and interobserver reliability was found for both tests.

Table 2 Reliability intra and inter-observer for the electroacoustic probe, Celaya, 2014 (n = 200)

The validity is shown for the three measurements for CST in Table 3; no significant differences were found. The sensitivity is low due to the bilateral affectations causing false negative results to be given.

Table 3 Validity of comparative test sound transmission, Celaya, 2014(n = 200)

The validity test for STE/F is shown in Table 4 for the three measurements. The sensitivity, specificity and predictive values are higher than with the CTST because each hip is evaluated separately diagnosing more accurately the bilateral cases; no significant differences were found between the three measurements for the different parameters of validity.

Table 4 Validity of sound transmission with extension/flexion, Celaya, 2014 (n = 200)

From the ultrasound employing the Graf method: for the left side, 85 healthy hips (Graf 1), 13 hips with physiological immaturity (Graf 2) and 2 hips with subluxation (Graf 3), were obtained; for right hip, 86 healthy hips (Graf 1) and 14 hips with physiological immaturity (Graf 2) were diagnosed.

Discussion

The sample was not representative of the infant population since the participation was by invitation and voluntarily, preventing the generalization of the results, which is a major drawback of the study.

There were no infants excluded as no rigid dislocations were detected.

The frequency of family history was high (16%) (Table 1). There may be a bias of the subject, for if they had relatives who had suffered DDH, they might have agreed to participate more easily, compared to those without such a history.

No diagnosis of complete hip dislocation was obtained.

Other disadvantages of the study is that only 2 subluxation cases were diagnosed and that physiologically immature hips can evolve mainly from healthy hips and/or from a small percentage subluxation.

Intraobserver and interobserver reliability of electroacoustic probe (Table 4) was higher than those reported by the bone radar [9].

The CTST shows a low sensitivity, because bilateral cases are detected as false negatives, from 37.9% to 44.8% in the three measurements; these are similar results to those reported by Padilla et., in 1996, of 27.27% with CTST in neonates using the tuning fork and stethoscope [6], and with the radar bone (Mexican Pat. N° 337,887) device a greater sensitivity of 60.9%9 was found because less bilateral cases were detected, but the specificity showed values ​​above 90%, a positive predictive value greater than 73%, and a negative predictive value greater than 90% (Table 3); similar results were reported with tuning fork and stethoscope [6, 8], and with the radar bone (Mexican Pat. N° 337,887) [9].

For the STE/F, validity was raised with percentages above 80% for sensitivity, and 90% for specificity and predictive values ​​(Table 4); similar but slightly lower results reported by Padilla et al. using the radar bone (Patent Pending, University of Guanajuato) [9] and similar to those reported by Padilla et al. using the tuning fork and stethoscope [6, 8].

Kwong et al., [12] designed a device for measuring the difference of sound transmission in the hip and found a sensitivity of 100% and specificity of 75% with cutoff points with a 2 dB difference. Celaya results show lower sensitivity (90%) and higher specificity (100%) (Tables 3 and 4).

The Primary Care Physicians can use the electroaccoustic device as inexpensive screening tool to detect alterations in the hip, and it complement physical exam of neonates.

Conclusions

The use of electroacoustic probe showed moderate sensitivity and high specificity, and high repeatability in immaturity hip.

It is needed further research in a population with more cases of DDH to demonstrate if the electroacoustic probe have higher sensitivity.

An advantage of the electroacoustic probe is that it detects from physiological immaturity to hip subluxation, as reported in the study.

Abbreviations

CTST:

Comparative test sound transmission

DDH:

Developmental dysplasia of the hip

STE/F:

Spund transmission with extension/flexion

References

  1. Cymet-Ramirez J, Alvarez-Martinez MM, Garcia-Pinto G, Frias-Austria R, Meza-Vernis A, Rosales-Muñoz ME, et al. El diagnóstico oportuno de la displasia de cadera. Enfermedad discapacitante de por vida. Consenso del Colegio Mexicano de Ortopedia y Traumatología. Acta Ortopedica Pediátrica. 2011;25(5):313–22.

    CAS  Google Scholar 

  2. Figueroa-Ferrari RC, Padilla-Raygoza N. Congenital dislocation of the hip in neonates macrosomic. Ultrasonographic aspects. Rev. Med. IMSS, 1994; 32(3):277–279.(Spanish).

  3. American Academy of Pediatrics Committee on quality improvement, Subcommittee on developmental dysplasia of the hip. Clinical practice guideline: early detection of developmental dysplasia of the hip. Pediatrics 2000;105(4):896–905.

  4. Fernández E. Congenital dislocation of the hip: reduction with Pavlik harness modified in children one year old. Rev Mex Ortop Traumatol 1989;3:30–34. (Spanish).

  5. Padilla N, Figueroa RC. Diagnosis of congenital hip dislocation through comparative sound transmission Rev Mex de Pediatr 1992: 59 149–151. (Spanish).

  6. Padilla N, Figueroa RC. Sound transmission tests in the diagnosis of congenital hip dislocation in the newbornRev Mex de Pediatr 1996; 63: 265–268. (Spanish).

  7. Stone MH, Richardson JB, Bennet JC. Another clinical test for congenital dislocation of the hip. Lancet 1987; 1:954–955. http://dx.doi.org/10.1016/S0140-6736(87)90296-0.

  8. Padilla-Raygoza N, Medina-Alvarez D, Ruiz-Paloalto ML, Cordova-Fraga T, Sosa-Aquino MA, Perez-Olivas AH. Diagnosis of developmental dysplasia of the hip using sound transmission in neonates. Health 2014; 6:2510–2516. http://dx.doi.org/10.4236/health.2014.618289

  9. Padilla-Raygoza N, Diaz-Guerrero R, Ruiz-Paloalto ML, Córdova-Fraga T, Sosa-Aquino MA, Perez-Oliva HA. Validity and reliability of a measuring device based on sound transmission for diagnosis of hip dysplasia in newborns. Adv Biosci Biotechnol, 2014;5(10):831–837. http://dx.doi.org/10.4236/abb.2014.510097.

  10. Arti H, Mehdinasab SA, Arti S. Comparing results of clinical versus ultrasonographic examination in developmental dysplasia of hip. J Res Med Sci. 2013;18(12):1051–5.

    PubMed  PubMed Central  Google Scholar 

  11. Kowalczyk B, Felus J, Kwinta P. Developmental dysplasia of the hip: the problems in the diagnosis process in our own experience. Med Wieku Rozwoj. 2005;9(3):395–406.

    PubMed  Google Scholar 

  12. Kwong KSC, Huang X, Cheng JCY, Evans JH. New technique for early screening of developmental dysplasia of the hip: pilot study. J Pediatr Orthop 2003; 23(3):347–351. http://dx.doi.org/10.1097/01241398-200305000-00013.

  13. Padilla-Raygoza N. LCCDatabase. 2016. Retrieved from https://osf.io/kpf5s/?view_only=0a9682c6e1c842ad8e1d9a66e8dcf038.

Download references

Acknowledgements

Thanks to Dr. Jaime Gonzalez Garcia for ultrasound diagnosis of the hip in newborns to Carolina Mendoza Lara for her help in the collection of clinical data and Miss Lizbeth for the English review.

The authors wish to thank the Directorate for Research Support and Postgraduate Programs at the University of Guanajuato for their support in the translation and editing of the English-language version of this article.

Funding

The author received a financial support for this research from Direction of Support for Research and Postgrade, University of Guanajuato, Mexico, and they did not participate in the study.

Availability of data and materials

The dataset supporting the conclusions of this article is available in the Open Science Framework [13] (https://osf.io/kpf5s/?view_only=0a9682c6e1c842ad8e1d9a66e8dcf038).

Authors’ contribution

NP-R, designed the protocol, generated the data based, made statistical analysis, and wrote the final report. GO-V obtained the clinical data, and participated in writing the final report

SCD-S, obtained the consent forms, reviewed the ultrasound diagnosis of the hip, and participated in writing the final report. TC-F, designed the electroacoustic probe and supervised its functioning. MAS-A, together with TC-F, designed the electroacoustic probe and train in its use. VB-C, participated in the design of the protocol and statistical analysis, reviewed and approved the final report.

All authors read and approved the final version of the article.

Competing interests

The authors declare no potential conflicts of interest respect to research, authorship, and/or publication of this research.

Consent for publication

Not applicable

Ethics approval and consent to participate

The protocol was reviewed and approved by Bioethics Committee of Division of Health Sciences and Engineering, Campus Celaya Salvatierra, University of Guanajuato Mexico with registry number CIDSIC-1501310

All parents of neonates signed the informed consent form.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Nicolas Padilla-Raygoza.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Padilla-Raygoza, N., Olvera-Villanueva, G., del Carmen Delgado-Sandoval, S. et al. Validity and reliability of electroacoustic probe for diagnosis of developmental dysplasia of the hip. BMC Pediatr 17, 149 (2017). https://doi.org/10.1186/s12887-017-0903-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12887-017-0903-z

Keywords