Skip to main content

Table 2 Summary of environmental exposure conditions and relevant standards

From: Avoid equipment graveyards: rigorous process to improve identification and procurement of effective, affordable, and usable newborn devices in low-resource hospital settings

Test condition

Description

Relevant standard

High temperature

Constant device operation under high heat; 50 °C and relative humidity < 50% for 16 h

IEC 60068–2-2

High humidity

Constant device operation under high humidity; relative humidity 95% cycling between 25 °C and 40 °C for 48 h, held at each temperature twice for 12 h

IEC 60068–2-30

Dust exposure

Recirculated 400 g/m2 dust particles < 75 μm for 4 h

IEC 60069–2-68

Line voltage fluctuations and power failure

Varied line voltage by ± 8%, ± 12%, and simulated total power failure

IEC 61000–4-14

  1. Abbreviations: h hours, IEC International Electrotechnical Commission