Table 1 Frequency of assessed variables during the rounds of syringe infusion pump audits

From: Smart pumps and random safety audits in a Neonatal Intensive Care Unit: a new challenge for patient safety

Alaris® CC Syringe Pumps (N measurements)

N

Appropriate Use % (Confidence Interval)

Period 1 (N 52)

Period 2 (N 160)

Period 1

Period 2

P value

Line Type

Central

35

117

67.31 (54.56–80.05)

73.13 (66.26–79.99)

0.5275

Pressure alarm

Programmed

Yes

48

160

92.31 (85.07–99.55)

100 (97.72–100)

0.0031

Programmed

Good

1

132

1.92 (1.81–5.65)

82.50 (76.61–88.39)

0.0001

VTI

Programmed

Yes

37

134

71.15 (58.84–83.46)

83.75 (78.03–89.47)

0.0725

Programmed

Good

26

134

50 (36.41–63.59)

83.75 (78.03–89.47)

0.0001

Infusion rate

Good

51

160

98.08 (94.34–100)

100 (97.72–100)

0.5529

Appropriate use

Yes

0

117

0 (0–6.85)

73.13 (66.26–79.99)

<0.0001

  1. VTI: volume to infuse in a certain time