Table 2 Results of the Base Case and Sensitivity Analyses

From: Cost effectiveness of a novel device for improving resuscitation of apneic newborns

Parameter Description

(Base Case Value)

Direction

Value

Cost per DALY Averted (Min, Max)

Base Case (N/A)

N/A

N/A

24.44 (4.80, 73.62)

Proportion of Time of Leaking with AIR** (0.700)

Upper Estimate

0.85

51.27 (10.68, 510.02)

Lower Estimate

0.05

7.75 (1.80, 16.59)

Hazard Ratio Mortality of Moderate to Severe Impairment (11.36)

Upper Estimate

20.00

23.39 (5.20, 61.75)

Lower Estimate

2.50

26.43 (5.31, 109.50)

Hazard Ratio Mortality of Mild Impairment (3.15)

Upper Estimate

10.00

25.21 (4.66, 85.00)

Lower Estimate

1.50

27.10 (4.59, 174.21)

Facility Births per Year

Upper Estimate

9000

11.76 (2.72, 36.81)

Lower Estimate

2250

48.26 (10.83, 170.01)

Cost of AIR Device

Upper Estimate

250

44.63 (9.26, 129.06)

Lower Estimate

100

20.43 (4.78, 72.92)

Probability of Apnea at Birth (0.0547)

Upper Estimate

0.109

11.68 (2.48, 28.51)

Lower Estimate

0.027

52.48 (8.00, 259.71)

  1. Key: DALY (Disability-Adjusted Life Year), AIR (Augmented Infant Resuscitator)
  2. *Costs are presented as a mean value and a lower (minimum) and upper (maximum) bound that reflect the range of uncertainty in the calibrated parameter sets
  3. **Defined as (t AIR / t NO AIR), where t is the time spent leaking. Ratios closer to 0 are favorable for the AIR, as less time is spent leaking. Ratios closer to 1 reflect decreased effectiveness of the AIR, as the time spent leaking with the device is comparable to the time leaking without its use