Table 3 Diagnostic Accuracy of PCT point-of-care in Identifying Infants with Bacterial Infections with 95% Confidence Intervals

From: Point-of-care testing for procalcitonin in identifying bacterial infections in young infants: a diagnostic accuracy study

 

PCT Cut-off 0.25 ng/ml

PCT Cut-off 0.5 ng/ml

PCT Cut-off 1.0 ng/ml

Invasive Bacterial Infection

All Bacterial Infection

Invasive Bacterial Infection

All Bacterial Infection

Invasive Bacterial Infection

All Bacterial Infection

Sensitivity

1.00 (0.40 to 1.00)

0.86 (0.56 to 0.97)

1.00 (0.40 to 1.00)

0.86 (0.56 to 0.97)

1.00 (0.40 to 1.00)

0.86 (0.56 to 0.97)

Specificity

0.72 (0.63 to 0.80)

0.77 (0.68 to 0.84)

0.91 (0.84 to 0.95)

0.97 (0.92 to 0.99)

0.92 (0.85 to 0.96)

0.98 (0.93 to 0.99)

Positive Predictive Value

0.11 (0.03 to 0.26)

0.32 (0.18 to 0.49)

0.27 (0.09 to 0.55)

0.80 (0.51 to 0.95)

0.29 (0.10 to 0.58)

0.86 (0.56 to 0.97)

Negative Predictive Value

1.00 (0.95 to 1.00)

0.98 (0.93 to 1.00)

1.00 (0.96 to 1.00)

0.98 (0.93 to 1.00)

1.00 (0.96 to 1.00)

0.98 (0.93 to 1.00)